The 1620 is a fully piezo resistive silicon pressure sensor for use in invasive blood pressure monitoring. The sensor is designed to be used with automated assembly equipment and can be dropped directly into a customer's disposable blood pressure housing. The sensor is designed to meet the requirements as described in AAMI BP22 specification.

    PRODUCT INTRODUCTION

    The 1620 is a fully piezo resistive silicon pressure sensor for use in invasive blood pressure monitoring. The sensor is designed to be used with automated assembly equipment and can be dropped directly into a customer's disposable blood pressure housing. The sensor is designed to meet the requirements as described in AAMI BP22 specification.

    The pressure sensor consists of a pressure sensing element mounted on a ceramic substrate. Thick film resistors on the ceramic substrate are laser trimmed for compensation and calibration. A plastic cap is attached to the ceramic substrate to provide an easy method of attachment to the customer’s assembly and protection for the sensing element. A dielectric gel is placed over the sensor to provide electrical and fluid isolation. 

    The products are shipped on a tape and reel. Performance characteristics and packaging can be easily tailored on a special order basis to meet the requirements of specific customers.

    Disposable IBP Transducers TE 1620's chip (6)s9u
    Disposable IBP Transducers TE 1620's chip (5)oa7
    Disposable IBP Transducers TE 1620's chip (7)6zp

    Advantages

    Model: 1620-H-W-T
    FEATURES APPLICATIONS
    Low Cost, Medical Applications Invasive Blood Pressure Monitoring
    Small Size and Reliable Performance Anesthesia department
    Gel Isolation for Liquids Interventional department
    Operates from 5°C to 45°C Intensive Care Unit
    Compatible for Automated Assembly Intracranial pressure
    1% Accuracy for Replacements Kidney Dialysis Machines
    5.0 uV/V/mmHg Sensitivity Urodynamic test
    OEM custom Intrauterine pressure measuring

    Notes

    1. Output of sensor with no pressure applied and a 150kΩ resistor placed across +SUPPLY to +OUTPUT. 

    2. Best fit straight line. 

    3. Defined as common mode symmetry between any output and supply terminal. 

    4. Over an 8 hour time period after a 10 second warm-up.

    5. Over operating temperature range 5°C–40°C with respect to 23°C. 

    6. One discharge per minute performed by customer. 

    7. Sterilization performed by customer, compatible with ETO, GAMMA, or E-Beam sterilization. 

    8. After curing, meniscus of gel shall be flush to .035” [.89mm] below surface

    9. This model is not suitable for surface mount operation.